Methodological issues of bioimpedance analysis: comparability of instruments and the impact of electrodes on measurement outcomes

Methodological issues of bioimpedance analysis: comparability of instruments and the impact of electrodes on measurement outcomes

Recieved: 07/23/2025

Accepted: 08/27/2025

Published: 02/18/2026

Keywords: biological anthropology; bioimpedance; bioimpedance instruments; bioimpedance examination technique; electrodes; body composition; phase angle

Available online: 18.02.2026

To cite this article

Rudnev Sergey G., Anisimova Anna V., Godina Elena Z., Zadorozhnaya Liudmila V. , Permiakova Ekaterina Yu. , Sipatrova Anastasya G., Khomyakova Irina A., Zubko Aleksandr V. Methodological issues of bioimpedance analysis: comparability of instruments and the impact of electrodes on measurement outcomes. // Lomonosov Journal of Anthropology 2026. Issue 1. 85-103 https://doi.org/10.55959/MSU2074-8132-26-1-7.

This work is licensed under a Creative Commons: Attribution 4.0 International (CC BY 4.0). (CC BY 4.0). (https://creativecommons.org/licenses/by/4.0/deed.ru)
Issue 1, 2026

Abstract

Introduction. The differences between bioimpedance instruments, electrodes, measurement techniques, and data processing algorithms raise questions about the consistency of bioimpedance data and body composition estimates. This study aims to compare the results of measurements taken with various bioimpedance devices and electrode types, including those used in Russian health centers.

Material and methods. Bioimpedance measurements were conducted on 20 adult volunteers, 10 women and 10 men, aged between 21 and 54 years. Four instruments were used in the study: ABC-01 "Medas" (SRC Medas, Moscow), Diamant-AIST (Diamant LLC, St. Petersburg), the prototype bioimpedance analyzer as part of the "Zdorovye-Express" hardware and software system with EK6C-03-"CARDi2/4" cardio amplifier (Medical Computer Systems LLC, Moscow, Zelenograd) and Tanita MC-780MA (Tanita, Japan). Eight types of disposable electrodes were used with the ABC-01 "Medas" instrument: Ambu White Sensor 0415M (Ambu, Denmark), Bianostic AT (Data Input, Germany), Eurotrode PFR2034 (Pirrone srl, Italy), F9049 / RU2234 TAB (FIAB, Italy), Schiller Biotabs 23x34 mm (Schiller, Switzerland), Skintact RT-34 (Leonhard Lang GmbH, Austria), TopTrace MedTab (Ceracarta, Italy), and 2100 Swaro-tab (Tyrolmed, Austria). Two types of reusable electrodes, manufactured by Diamant LLC and Medical Computer Systems (MCS) LLC, were used with the Diamant-AIST and Zdorovye-Express instruments. Electric properties of the bioadhesive electrodes were assessed using a sandwich test. A comparison of the data obtained from volunteers using different bioimpedance instruments and electrodes was conducted using the Wilcoxon signed rank test for paired observations, as well as one-way analysis of variance, with a significance level set at p=0.05.

Results and discussion. Our study showed the presence of significant inter-instrument differences in bioelectrical parameters and body composition estimates. The maximum difference in the average value of fat-free mass and body fat between the ABC-01 "Medas", Diamant-AIST and Tanita MC-780MA instruments was 2.8 kg for men and 2.2 kg for women, whereas for the percentage body fat it was 3.9% for men and 3.7% for women. There were also significant deviations in body composition estimates obtained with the Zdorovye-Express analyzer compared to estimates from the other instruments. The BIA measurements data using the ABC-01 "Medas" instrument and the types of disposable bioadhesive electrodes commonly used in Russian Health Centers were in good agreement, despite the wide variation in electrical properties of the electrodes (their intrinsic impedance ranged from 96.8 ohm for 2100 Swaro-tab to 694.3 ohm for Eurotrode PFR2034). However, compared to the reference Bianostic AT electrodes, which had an intrinsic impedance of 18.9 ohm, the subgroups of women and men showed an average overestimation of 0.15 degrees in phase angle, while the subgroup of men also showed an overestimation in fat-free mass and an underestimation in fat mass, averaging between 0.3 and 0.4 kilograms. When comparing the reusable electrodes from Diamant LLC and MCS LLC, there were no significant differences in the bioelectrical parameters or body composition estimates in the study group.

Conclusion. To directly compare the results of measurements from different BIA instruments, their preliminary mutual calibration is necessary. The bioimpedance data obtained in Russian Health Centers using the ABC-01 "Medas" instrument and the specified types of disposable bioadhesive electrodes can be combined and analyzed together subject to the measurement procedure being followed. The reusable electrodes for bioimpedance measurements produced by Diamant LLC and Medical Computer Systems LLC are effectively interchangeable.

Funding. The study was performed at the Federal State Budgetary Institution "RIH" (“Russian research Institute of Health”) of the Ministry of Health of the Russian Federation within the framework of the Russian Science Foundation project No. 20-15-00386 (directed by V.I. Starodubov).

Acknowledgements. The study was conducted under the state assignment of Lomonosov Moscow State University (Godina E.Z., Zadorozhnaya L.V., Khomyakova I.A., Permiakova E.Yu.)The authors would like to thank the management and staff of Medical Computer Systems LLC (Zelenograd), and personally I.S. Reshetnikov and D.A. Prilutsky, for their help in organizing and conducting the study.

References

Bennett J.P., Cataldi D., Liu Y.E., Kelly N.N., Quon B.K. et al. Variations in bioelectrical impedance devices impact raw measures comparisons and subsequent prediction of body composition using recommended estimation equations. Clin. Nutr. ESPEN, 2024, 63, pp. 540–550. https://doi.org/0.1016/j.clnesp.2024.07.009

Bioelectrical impedance analysis in body composition measurement: National Institutes of Health Technology Assessment Conference Statement. Am. J. Clin. Nutr., 1996, 64(3 Suppl.), pp. 524S–532S. https://doi.org/10.1093/ajcn/64.3.524S

Boone K.G., Holder D.S. Effect of skin impedance on image quality and variability in electrical impedance tomography: a model study. Med. Biol. Eng. Comput., 1996, 34 (5), pp. 351–354. https://doi.org/10.1007/BF02520003

Bunak V.V. Antropometry. Moscow, Uchpedgiz Publ., 1941. 368 p. (In Russ.).

Carter J.E.L., Heath B.H. Somatotyping: development and applications. Cambridge: Cambridge University Press, 1990. 520 p.

Chumlea W.C., Guo S.S., Kuczmarski R.J., Flegal K.M., Johnson C.L. et al. Body composition estimates from NHANES III bioelectrical impedance data. Int. J. Obes., 2002, 26 (12), pp. 1596-1609. https://doi.org/0.1038/sj.ijo.0802167

Dupertuis Y.M., Pereira A.G., Karsegard V.L., Hemmer A., Biolley E. et al. Influence of the type of electrodes in the assessment of body composition by bioelectrical impedance analysis in the supine position. Clin. Nutr., 2022, 41 (11), pp. 2455–2463. https://doi.org/10.1016/j.clnu.2022.09.008

Kushner R.F., Schoeller D.A. Estimation of total body water by bioelectrical impedance analysis. Am. J. Clin. Nutr., 1986, 44 (3), pp. 417–424. https://doi.org/10.1093/ajcn/44.3.417

Kyle U.G., Bosaeus I., De Lorenzo A.D., Deurenberg P., Elia M. et al. Bioelectrical impedance analysis-part II: utilization in clinical practice. Clin. Nutr., 2004, 23 (6), pp. 1430–1453. https://doi.org/10.1016/j.clnu.2004.09.012

Negasheva M.A. Anthropometry basics. Moscow, Ekon-Inform Publ., 2017, 216 p. (In Russ.).

Nescolarde L., Lukaski H., De Lorenzo A., de-Mateo-Silleras B., Redondo-Del-Río M.P., Camina-Martín M.A. Different displacement of bioimpedance vector due to Ag/AgCl electrode effect. Eur. J. Clin. Nutr., 2016, 70 (12), pp. 1401–1407. https://doi.org/10.1038/ejcn.2016.121

Permiakova E.Yu., Sipatrova A.G., Godina E.Z., Anisimova A.V., Zadorozhnaya L.V. et al. On the quality of skinfold measurements using skinfold and sliding calipers. Vestnik Moskovskogo universiteta. Seriya XXIII. Antropologiya, 2021, 2, pp. 5–25. (In Russ.). https://doi.org/10.32521/2074-8132.2021.2.005-020

Roos A.N., Westendorp R.G.J., Froehlich M. Tetrapolar body impedance is influenced by body posture and plasma sodium concentration. Eur. J. Clin. Nutr., 1991, 46 (1), pp. 53–60.

Rudnev S., Burns J.S., Williams P.L., Lee M.M., Korrick S.A. et al. Comparison of bioimpedance body composition in young adults in the Russian Children's Study. Clin. Nutr. ESPEN, 2020, 35 (1), pp. 153–161. https://doi.org/10.1016/j.clnesp.2019.10.007

Rudnev S.G., Starunova O.A., Godina E.Z., Ivanova A.E., Zubko A.V., Starodubov V.I. The Russian bioimpedance database: an update. J. Electr. Bioimp., 2022, 13, pp. 66–72. https://doi.org/10.2478/joeb-2022-0010

Rush E.C., Crowley J., Freitas I.F., Luke A. Validity of hand-to-foot measurement of bioimpedance: standing compared with lying position. Obesity, 2006, 14 (2), pp. 252–257. https://doi.org/10.1038/oby.2006.32

Silva A.M., Campa F., Stagi S., Gobbo L.A., Buffa R. et al. The bioelectrical impedance analysis (BIA) international database: aims, scope, and call for data. Eur. J. Clin. Nutr., 2023, 77 (12), pp. 1143–1150. https://doi.org/10.1038/s41430-023-01310-x

Sipatrova A.G., Godina E.Z., Permiakova E.Yu., Anisimova A.V., Zubko A.V., Rudnev S.G. Bioimpedance assessment of body composition using ABC-01 "Medas" and Diamant-AIST instruments: a comparison. Lomonosov Journal of Anthropology (Vestnik Moskovskogo Universiteta. Seriya XXIII. Antropologiya), 2023, 2, pp. 70–81. (In Russ.). https://doi.org/10.32521/2074-8132.2023.2.070-081

Smirnov A.V., Kolesnikov V.A., Nikolaev D.V., Eryukova T.A. ABC-01 "Medas": Analyzer for the Assessment of Body Fluids Balance with Software (User Manual). Moscow, SRC Medas Publ., 2009. 38 p. (In Russ.).

Stratton M.T., Smith R.W., Harty P.S., Rodriguez C., Johnson B.A. et al. Longitudinal agreement of four bioimpedance analyzers for detecting changes in raw bioimpedance during purposeful weight gain with resistance training. Eur. J. Clin. Nutr., 2021, 75 (7), pp. 1060–1068. https://doi.org/10.1038/s41430-020-00811-3

Tinsley G.M., Moore M.L., Silva A.M., Sardinha L.B. Cross-sectional and longitudinal agreement between two multifrequency bioimpedance devices for resistance, reactance, and phase angle values. Eur. J. Clin. Nutr., 2020, 74 (6), pp. 900–911. https://doi.org/10.1038/s41430-019-0496-8

Ward L.C. Inter-instrument comparison of bioimpedance spectroscopic analysers. Open Med. Devices J., 2009, 1, pp. 3–10. https://doi.org/10.2174/1875181400901010003